Related information
> 【News Release】Practical application of simple and rapid E. coli measurement technology 2020. Sep 23
published in the Japan Sewage Works Newspaper (Nihon-gesuido-shimbun). 2020.Sep 16
Specifications
Measurement method: Enzyme activity assay
Excitation wavelength: 360 nm
Measurement wavelength: 460 nm
Samples: Sewage, industrial wastewater, river water, groundwater, drinking water
Species reactivity: E.coli
Storage temperature: Store at 2-8 ℃
Expiration date: 6 months after production
Measuring equipment: Fluorescence microplate reader
Kit contents
The plate with E. coli selective freeze-dried medium 96 well × 4
Features
・Measure 96 samples at the same time
・Measurement time is only 3 hours
・Reasonable price for 250 JPY per test
・Only 200 μL of sample volume is needed
Operation procedure
(1) Set the plate reader at 37 °C

(2) Make holes in the seal attached to the plate (or remove the seal.)

(3) Dispense 200μL of the test sample into each well of medium-containing 96-well plate.

(4) Set the excitation wavelength at 360 nm and the measurement wavelength at 460 nm. Measure every 10 minutes for 3 hrs.
Example
.png)